XL2 Audio and Acoustic Analyzer
Sound Level Meter
| Product Configurations |
| • |
XL2 + M2210: |
• Class 1 frequency response according IEC 61672, IEC 60651
• Type 1 frequency response according ANSI S1.4 |
| • |
XL2 + M4260: |
• Class 2 according IEC 61672, IEC 60651
• Type 2 according ANSI S1.4 |
|
| Conforms with Standards |
| • |
IEC 61672, IEC 60651, IEC 60804, IEC 61260 class 0, ISO 2969 |
| • |
China: GB/T 3785:2010, GB/T 3241, GB 3096-2008, GB 50526, GB-T_4959-1995 |
| • |
Germany: DIN 15905-5, DIN 45645-2, optional: DIN 45645-1 |
| • |
Japan: JIS C1509-1:2005, JIS C 1513 class 1, JIS C 1514 class 0 |
| • |
Switzerland: SLV 2007 |
| • |
US: ANSI S1.4, ANSI S1.43, ANSI S1.11-2004 class 1 |
| • |
International IEC standards are adopted as European standards and the letters IEC are replaced by EN. XL2 conforms to these EN standards. |
|
| Measurements |
| • |
SPL actual, Leq, Lmin, Lmax, LCpeak |
| • |
Gliding LAeq with t = 5”, 10’, 15’ and 60’ |
| • |
All measurement results are simultaneously available |
| • |
Correction value measurement wizard |
| • |
Noise curves NC, NR, PNC, RC with post processing |
| • |
Noise exposure level LEX with post processing |
| • |
Logging all data or subsets in selectable intervals |
| • |
Recording of wav-files (ADPCM), a new wav-file starts every 12 hours (max. wav-file size 512 MB) |
| • |
Recording of voice notes |
| • |
Limit monitoring showing exceeding sound levels |
| • |
Digital I/O interface for control external peripherals |
| • |
Level Resolution: 0.1 dB |
| • |
Internal noise: 1.3 µV A-Weighted |
|
|
Weighting
|
| • |
Frequency weighting: A, C, Z |
| • |
Time weighting: Fast, Slow, optional: Impulse |
| • |
Measurement Bandwidth (-3dB): 2 Hz to 23.6 kHz |
|
| Real Time Analyzer |
| • |
Frequency Resolution: 1/3 octave, 1/1 octave, wide band |
| • |
Frequency Range: 6.3 Hz to 20 kHz |
| • |
Level Resolution: 0.1 dB |
| • |
Band pass filters conform IEC 61260 class 0, ANSI S1.11-2004, class 1 |
| • |
Frequency weighting: X-Curve acc. ISO 2969 |
| • |
Capturing of single reading into the internal memory for comparative measurements |
|
Options for Sound Level Meter and Acoustic Analyzer
Remote Measurement
(optional) |
| • |
Querying measurement data online via the USB interface |
|
|
Functions of
Extended Acoustic
Pack (optional)
|
| • |
SLMeter/RTA function
| - |
Percentile statistics for wide band, 1/1 and 1/3 octave spectrum:
. 1%, 5%, 10%, 50%, 90%, 95%, 99%
. Sampling: every 1.3 ms
. Wide band: in 0.1 dB wide classes, based on sampling LAF
. 1/1 and 1/3 octave spectrum: in 1.0 dB wide classes,
based on Lxy (x= A, C or Z, y= F or S)
. Dynamic range: 140 dB
|
| - |
Time weighting: Impulse |
| - |
Sound Pressure Level LIeq |
| - |
Sound Exposure Level LAE |
| - |
Event-triggered audio and data recording |
| - |
Recording of wav-files (24 bit, 48 kHz), a new wav-file is started after each hour (max. wav-file size 512 MB) |
| - |
True peak level in 1/1 and 1/3 octave resolution |
| - |
Clock-impulse maximum level (TaktMax) and values as specified in DIN 45645-1 |
|
| • |
FFT function
| - |
High-resolution Zoom-FFT with selectable frequency ranges and resolution up to 0.4 Hz in the range of 5 Hz to 20 kHz |
|
| • |
RT60 function
| - |
Reverberation time RT60 in 1/3 octave resolution |
|
|
|
Functions of
Spectral Limits
Option (optional)
|
| • |
SLMeter/RTA function
| - |
True peak level in 1/1 and 1/3 octave resolution |
|
| • |
FFT function
| - |
High-resolution Zoom-FFT with selectable frequency ranges and resolution up to 0.4 Hz in the range of 5 Hz to 20 kHz |
|
| • |
1/12 octave function
| - |
High resolution RTA function “1/12 Oct + Tol” |
| - |
Selectable 1/1, 1/3, 1/6 and 1/12 octave spectral resolution |
|
| • |
FFT and 1/12 octave function
| - |
Capturing of multiple readings into the internal memory |
| - |
Comparing measurement results against captures with relative or absolute curve display |
| - |
Comprehensive tolerance handling |
| - |
Creating tolerance masks based on captures for passed/failed measurements |
| - |
Export and import of tolerance and capture files |
|
|
Acoustic Analyzer
| FFT Analysis |
| • |
Real time FFT with actual level, Leq, Lmin, Lmax |
| • |
Level Resolution: 0.1 dB |
| • |
Frequency ranges: 7 Hz - 215 Hz, 58 Hz - 1.72 kHz, 484 Hz - 20.5 kHz with 142 frequency bins shown on display |
| • |
Optional: High-resolution FFT, up to 0.4 Hz resolution in the range of 5 Hz - 20 kHz
|
| • |
Optional: Capture and tolerance function with multiple readings for comparative measurements and passed/failed analysis
|
|
| Reverberation Time RT60 |
| • |
1/1 octave bands results from 63 Hz - 8 kHz, based on T20 |
| • |
Optional: 1/3 octave bands results from 50 Hz - 10 kHz, based on T20 |
| • |
Range: 10 ms - 14 seconds |
| • |
Measurement according to ISO 3382 by Schroeder-method |
| • |
Test signal: Impulse source or gated pink noise generated by the
MR-PRO, MR2 or the included NTi Audio Test CD |
|
| Delay Time |
| • |
Propagation delay between electrical reference signal and acoustical signal using the internal microphone |
| • |
Resolution: 0.1 ms |
| • |
Range: 0 ms - 1 second (0 m - 344 m) |
| • |
Test signal: NTi Audio delay test signal generated by the MR-PRO, MR2 or the included NTi Audio Test CD |
|
| Polarity |
| • |
Checks polarity of speakers and line signals |
| • |
Positive/Negative detection of wideband and individual 1/1 octave bands through internal microphone or XLR/RCA connector |
| • |
Test signal: NTi Audio polarity test signal generated by the MR-PRO, MR2 or the included NTi Audio Test CD |
|
| 1/12 Octave Analyzer (optional) |
| • |
Actual level, Leq, Lmin, Lmax |
| • |
Selectable 1/1, 1/3, 1/6 and 1/12 octave spectral resolution |
| • |
Capturing of multiple readings into the internal memory |
| • |
Comparing measurement results against captures with relative or absolute curve display |
| • |
Comprehensive tolerance handling |
| • |
Creating tolerance masks based on captures for passed/failed measurements |
|
STI-PA
Speech Intelligibility
(optional) |
| • |
Single value STI and CIS test result according to IEC 60268-16 (2003), ISO 7240-16 and ISO 7240-19 (replacing IEC 60849), DIN VDE 0828-1, DIN VDE 0833-4 |
| • |
Modulation indices and individual band level results with error indicator, post processing with recorded spectra supported |
| • |
Test signal: NTi Audio STI-PA signal generated by the MR-PRO, NTi Audio TalkBox or the STI-PA Test CD |
|
Audio Analyzer
| Level RMS |
| • |
True RMS detection in V, dBu, dBV and dBSPL |
| • |
Range XLR/RCA input: 2 µV - 25 V (-112 dBu to +30 dBu) |
| • |
Accuracy: ± 0.5 % @ 1 kHz |
| • |
Flatness: ± 0.1 dB @ 12 Hz to 21.3 kHz |
| • |
Bandwidth (-3dB): 5 Hz to 23.6 kHz |
| • |
Resolution: 3 digits (dB scale) or 4 digits (linear scale) |
|
| Frequency |
| • |
Range: 9 Hz to 21.3 kHz |
| • |
Resolution: 6 digits |
| • |
Accuracy: < ± 0.003% |
|
THD+N
(Total Harmonic
Distortion + Noise) |
| • |
Range: -100 dB to 0 dB (0.001% to 100%) |
| • |
Minimum level: > -90 dBu |
| • |
Fundamental frequency range: 10 Hz to 21.3 kHz |
| • |
Measurement bandwidth: 5 Hz to 23.6 kHz |
| • |
Resolution: 3 digits (dB scale), 5 digits (linear scale) or 6 digits (x1 scale)
|
| • |
Residual THD+N @ XLR/RCA input: < 2 µV |
|
| Scope |
Auto ranging, auto scaling |
| Filter |
Frequency weighting: A, C, Z
Highpass 100 Hz, 400 Hz, 19 kHz
Bandpass 22.4 Hz - 22.4 kHz acc. IEC468-4
|
Remote Measurement
(optional) |
| • |
Querying measurement data online via the USB interface |
|
Input / Output Interfaces
| Audio Inputs |
| • |
XLR balanced with input impedance = 200 kOhm, phantom power: +48 V switchable, automated sensor detection for M4260/M2210 microphones |
| • |
RCA unbalanced with input impedance >30 kOhm |
| • |
Built-in condenser microphone for polarity testing, delay measurements and voice note recording |
|
| Audio Outputs |
| • |
Built-in speaker |
| • |
Headphone connector 3.5 mm Minijack Stereo |
|
| USB Interface |
USB mini connector for data transfer to PC and charging of Li-Po battery |
| Serial I/O |
Optional 1-wire interface for control of external peripherals |
| TOSLINK |
24 bit linear PCM audio signal output (optional, prepared for later firmware extension)
|
| Memory |
Mini-SD card included (2 GByte), removable, storing measurement data in ASCII format, screen shots, voice notes and wav-files
|
| Power Supply |
| • |
Rechargeable Li-Po battery included, type 3.7 V / 2260 mAh,
typical battery lifetime 4 hours |
| • |
Dry cell batteries type AA, 4 x 1.5 V, typical battery lifetime 4 hours
|
| • |
Linear external power supply 9 VDC (charges Li-Po battery in operation) |
| • |
Range: 7.5 - 23.0 VDC @ minimum 6 Watt |
|
General
| Clock |
Real time clock with lithium backup battery
|
| Calibration |
| • |
Recommend calibration interval: one year
|
| • |
Microphone calibration supported with external calibrator |
| • |
Calibration certificate for new instruments optional available |
|
| Mechanics |
| • |
Tripod mount and wire stand mounted on rear side
|
| • |
Display: 160 x 160 pixels grey scale with LED back light |
| • |
Dimensions: 180 mm x 90 mm x 45 mm (7.1” x 3.5” x 1.8”) |
| • |
Weight: 480 g (1 lbs) including built-in Li-Po battery |
|
| Temperature Range |
-10 °C to +50 °C (14° to 122°F) |
| Humidity |
5% to 90% RH, non condensing
|
Electromagnetic
Compatibility |
CE compliant: EN 61326-1 Class B, EN 55011 class B, EN 61000-4-2 to -6 and -11
|
All information are subject to change without notice.
|
| |
|
|
| |
|
M4260 |
M2210
Extended Performance |
| Microphone Type |
|
Omni-directional, pre-polarized condenser,
free-field microphone
|
| Capsule / Transducer |
|
1/4”
permanently installed capsule
|
1/2” detachable capsule with
60UNS2 thread
|
| PreAmplfier Type |
|
- |
MA220 |
| Flatness |
|
Class 2 (IEC61672-1)
+1/-4.5 dB @ 5 Hz - 20 Hz
±1.5 dB @ 20 Hz - 4 kHz
±4.5 dB @ 4 kHz - 20 kHz
|
Class 1 (IEC61672-1)
±1 dB @ 5 Hz - 20 Hz
±1 dB @ 20 Hz - 4 kHz
±2 dB @ 4 kHz - 20 kHz
|
| Frequency Response |
|
available as MS Excel-data on request |
| Frequency Range |
|
5 Hz - 20 kHz |
5 Hz - 20 kHz |
| Residual Noise Floor typical |
|
29 dB(A) |
19 dB(A) |
Maximum SPL
@ THD 3%, 1 kHz |
|
144 dBSPL |
144 dBSPL |
| Sensitivity typical |
|
-31.7 dBV/Pa ±3 dB
(26 mV/Pa) @ 1kHz |
-34 dBV/Pa ±3 dB
(20 mV/Pa) @ 1kHz |
| Temperature Coefficient |
|
±0.02 dB / °C @ -10°C to +50°C
(14°F to 122°F) |
±0.015 dB / °C @ -10°C to +50°C
(14°F to 122°F) |
| Pressure Coefficient |
|
not defined |
-0.02 dB / kPa |
| Influence of Humidity |
|
not defined |
< ±0.01 dB (non condensing) |
| Long Term Stability |
|
not defined |
> 250 years / dB |
| Temperature Range |
|
-10°C to +50°C (14°F to 122°F) |
| Humidity |
|
5% to 90% RH, non condensing |
| Electronic Data Sheet |
|
NTi Audio ASD according IEEE P1451.4 V1.0, Class 2, Template 27 |
| Output Impedance |
|
100 Ohm balanced |
| Power Supply |
|
48 VDC phantom power, 3 mA typical |
| Dimensions |
|
Length 150 mm (5.9”), diameter 20.5 mm (0.8”) |
| Connector |
|
3-pole XLR (NEUTRIK) |
| Weight |
|
83 g, 2.93 oz |
100 g, 3.53 oz |
| NTi Audio # |
|
600 040 025 |
600 040 020 |
Typical Frequency Response*

* NTi Audio defines tighter tolerances lines than specified in the standard IEC 61672. These tolerances are shown as thicker grey lines.